|
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and visually observed the sorter stopped in the z-axis position.The fse also observed the sorter z-axis bearings and identified lubrication was needed.Fse lubricated the sorter z-axis bearings, performed cup macro three times and was within specifications.Fse validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for failure mode "4055 sorter z-axis motor slippage detected " for the aia-2000 analyzer from (b)(6) 2025 through aware date 26mar2026.There were two (2) similar complaints identified during the search period, including this case.A 13-month complaint history review and service history review for similar complaints was performed for failure mode "2070 clogging detected during specimen suction by main arm" for the aia-2000 analyzer from (b)(6) 2025 through aware date 26mar2026 there were nineteen (19) similar complaints identified during the search period, including this case.A 13-month complaint history review and service history review for similar complaints was performed for failure mode "noise" for the aia-2000 analyzer from (b)(6) 2025 through aware date 26mar2026 there were twenty (20) similar complaints identified during the search period, including this case.Escalation review a 13 month retrospective review revealed revealed nineteen (19) occurrences of 2070 clogging detected during specimen suction by main arm" and twenty (20) occurrences of audible noise on the aia-2000 analyzer.Causes usually are associated with mechanical damage, sample processing, mishandling, misalignment, clogging/blockage or wear and tear.Data assessment determined the occurrences were mostly attributed to obstruction of flow/maintenance problem.All causes of the reported event were corrected prior to reporting patient results.Regular analyzer maintenance and following proper procedure will minimize the reported errors.There is no indication of a systemic issue and there is no impact to patient safety.Aia-2000 operator's manual in appendix 4: error messages states the following: (4055) sorter z-axis motor slippage detected cause: the each-limit sensor detected slippage after sorter z-axis movement.Solution: contact tosoh service center or local representatives.(2070) clogging detected during specimen suction by main arm cause: the negative pressure detected after specimen suction exceeded the standard.The specified amount of specimen may not have been obtained because the sampling nozzle was blocked.The measurement result will be flagged (sc flag).Solution: verify that the specimen is free of solid substances (such as fibrin) or that there is sufficient volume of specimen if it is prepared in the primary tube and retry the measurement.If retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event was due to z-axis bearings of the sorter needed lubrication.
|