• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 56MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 56MM Back to Search Results
Catalog Number 940-02-56H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/01/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The agent reported "(lesion behind cup, changed acetabular components)".The previous surgery and the surgery detailed in this event occurred 3 years and 3 months apart.Item number: 940-02-56h, item description: empowr acet system, cup, hemispherical, cluster hole, 56mm, lot number: 802z1162, part revision: c, product type: hip, manufacture date: 11/02/22, expiration date: 10/22/28.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to lesion.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.
 
Event or Problem Description
Revision surgery - due to lesion behind cup, changed acetabular components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR ACET SYSTEM, CUP, HEMISPHERICAL, CLUSTER HOLE, 56MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key24974723
Report Number1644408-2026-00747
Device Sequence Number4979111
Product Code LPH
UDI-Device Identifier00190446311584
UDI-Public00190446311584
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number940-02-56H
Device Lot Number802Z1162
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2026
Initial Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-282 LOT: 858B1296; 940-00-030 LOT: 147Z1132; 951-01-46H LOT: 2219A1032; 952-28-46H LOT: 2227A1045
Outcome Attributed to Adverse Event Hospitalization;
Patient Age66 YR
Patient SexMale
-
-