| Model Number |
PM3262 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No complaint was received with the return of the device.Failure event observed during analysis.The device was received with normal telemetry and output.Review of the field settings indicates auto-capture was on.When automatic settings are programmed, such as auto-capture, the device output would adjust itself to accommodate the patient¿s needs for pacing.Visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Longevity assessment found the device was operating at the elective replacement indicator (eri) voltage consistent with normal usage.A manufacturing process anomaly consistent with header bonding anomaly may have occurred.
|
| |
|
Search Alerts/Recalls
|