The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the ccd drive pcb fluid damage, the ultrasound connector body fluid damage, the lg cable connector fluid damage, the lg cable connector corroded, the shield cover corroded, the insertion flexible tube crushed, the operation channel (primary) resistance, and the angulation down angulation decrease; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|