|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module with drive pcb.In addition, our technician confirmed that the nozzle gluing missing, and the objective lens dirty; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|