| Catalog Number |
10487040 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported falsely depressed innovance vwf ac (von-willebrand factor activity) results were obtained on two patient samples on cs-5100 instrument.The erroneous results were reported to the physician(s).The samples were repeated on the same instrument.The repeat results were higher than the erroneous results.The repeat results were reported, as the correct results, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, depressed vwf-ac results.
|
| |
|
Additional Manufacturer Narrative
|
|
An outside of the united states (ous) customer contacted siemens customer care center to report discordant results that were obtained on two patient sample on the cs-5100 instrument for the assay innovance vwf ac.Siemens is investigating the issue.In addition the customer reported a discordant result that was obtained for one patient sample on the cs-5100 instrument for the assay vwf ag.A separate report was filed for vwf ag.Innovance vwf ac reagent is marketed in the united states under siemens material number 10873906.The 510(k) number documented in section g4 is for the us product.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 9610806-2026-00009 on 24-apr-2026.Additional information received 18-may-2026: siemens reviewed the information provided and data obtained.Neither the calibration curve nor the control charts showed any anomalies during the period when the initial results were obtained.According to technical service, system was calibrated on december 02, 2025 and march 03, 2026.Quality controls were always in range.Investigation of the received back-up data did not produce any additional knowledge referring to circumstances of the analysis of the patient samples affected.Re-testing of affected samples was done after service took place.Review of service report showed that a leakage at the syringe of sample arm a was identified.The complete syringe was subsequently replaced.To check the system performance after service visit, control run was performed.In addition, two tests each with five patient samples were measured with vwf in micro mode and normal mode.Primary tubes were placed capped and uncapped.The customer has resumed routine operation with the system after repair.The issue was solved by routine troubleshooting, replacing a spare part.No systematic hardware or reagent problem was identified.Based on the data evaluated and the complaint description, there is no indication that further results are affected.After service/repair system is operational again.Sample arm a was exchanged.Customer is operational.In section h6, the investigation findings and investigation conclusions codes were updated.The reported behavior was resolved by standard service actions/routine troubleshooting.The instrument is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Search Alerts/Recalls
|