|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section d1 (brand name), d4 (model #), and d4 (primary udi#).The device was returned to zoll medical corporation.The customer's report was observed in the log but could not be reproduced.After clearing the error, the device had 30 sec self-tests run (runs a daily, weekly, and monthly self-test within 30 seconds) without any discrepancies noted.On/off current, patient and circuit impedance testing, internal inspection, and shock testing all passed.The pads used during the reported event were not returned for evaluation.The main board was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|