Investigation evaluation: the product said to be involved was returned in a white plastic bag.Provided with the return was an open box from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The distal catheter's wire guide lumen has split proximally from the catheter's distal tip, creating a burr.The cutting wire responded and bowed with handle manipulation confirming the cutting wire remains intact.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed in the photo provided.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the distal tip was split at the wire guide lumen.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.This device is packaged with a precurved stylet wire inside the wire guide lumen.If the handle is manipulated with the catheter in a coiled position or with the precurved stylet inside the cannulating tip, this could contribute to damage to the wire guide lumen.The instructions for use contain the following precaution: "do not exercise the handle while the device is coiled or the precurved stylet is in place, as this may cause damage to the sphincterotome and render it inoperable." the instructions for use advise the user with the following statement: ¿upon removing the device from the package, uncoil and straighten sphincterotome.Carefully remove the precurved stylet from the cannulating tip.¿ prior to distribution, all fusion omni-tome sphincterotomes are subjected to a visual inspection to ensure device integrity.The visual inspection verifies the tip is round with no sharp edges and no splitting.A review of the device history record confirmed that this lot met manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
|
During an endoscopic procedure, the physician used a cook fusion omni-tome.It was initially reported that the cutting wire tip was damaged.There was no reportable information at this time.Per our evaluation of the returned device on (b)(6) 2026, the tip of the catheter/wire guide lumen is damaged and there is a burr sticking out.[subject of report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|