| Model Number |
HARMONY-25 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10.Continuation - concomitant medical devices in use during the event include: harmony-dcs: product serial/lot number: (unknown); product type: (0195 - heart valves); implant date: not-implantable; explant date: na.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 days following the implant of a transcatheter pulmonary valve, a second transcatheter pulmonary valve was implanted valve-in-valve for an unspecified reason.
|
| |
|
Search Alerts/Recalls
|