• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number CNWTT0
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A health care professional reported that the intraocular lens (iol) was mounted according to the standard technique using the company's cartridge and injector.However, resistance encountered during delivery, resulting in apical amputation of the lens.Because the lens had already entered the eye, it was necessary to cut and remove it during the same procedure.Based on additional information received, the surgery performed was cataract phacoemulsification with intraocular lens implantation.The procedure involved patient's right eye.There was no patient harm, and the current patient condition was reported to be very well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24982245
Report Number1119421-2026-00878
Device Sequence Number21455357
Product Code MFK
UDI-Device Identifier00380652408233
UDI-Public(01)00380652408233(17)290422(11)250424(10)16022063
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNWTT0
Device Catalogue NumberCNWTT0.255
Device Lot Number16022063
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/31/2026
Supplement Date Received by Manufacturer05/12/2026
Initial Report FDA Received Date04/24/2026
Supplement Report FDA Received Date06/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1.8% HYALURONATE; CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C
Patient Age39 YR
Patient SexFemale
-
-