MEDOS INTERNATIONAL SÃ RL HD LAPSCP,EP,5.5,30,300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
242633 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Moisture Damage (1405)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer.Outer tube damaged.Illumination dispal tip fiber damaged.Optical system loose.Cosmetic issue scratches/dents rust/discoloration.Visual: deformed/bent, broken (2+ pieces): chipped/shaved/delaminated, visual: scratched/nicked.Per service report, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
It was reported that during in-house engineering evaluation, it was determined that the hd lapscp,ep,5.5,30,300 device was chipped/shaved/delaminated.It was initially reported that during a laparoscopy surgical procedure, it was discovered that multiple devices did not function to the doctor's expectations; they swapped devices, but issues persisted.The surgery was completed, and the next surgery was cancelled due to these issues.Camera images were blurry and could not be focused across the entire field of view.Light from the light cords was dim.The couplers were not focusing and there appeared to be fluid ingress in the scopes.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|