| Catalog Number |
795231 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the control panel arm on an epiq cvx ultrasound system dropped rapidly.This issue was found outside of clinical use, and there was no patient or user harm.The service engineer replaced the control panel arm gas struts to resolve the customer¿s immediate issue.Philips has started an investigation, and upon completion, a follow-up report will be submitted.
|
| |
|