| Catalog Number |
A1059 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - inspection of the skull clamp shows that the lock has rotational and lateral movement and a residue buildup is present.To resolve the issues, new components were added to replace worn internal parts and general cleaning, and maintenance were performed.Root cause - the complaint is confirmed via inspection of the unit.The unit needed cleaning and replacement of worn parts.The probable root cause is routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|
| |
|
Event or Problem Description
|
|
A facility reported that the c lock on the mayfield modified skull clamp (a1059) is sticking, hard to loosen, and when locked on the patient¿s head is hard to unlock.Additional information has been requested.
|
| |
|