| Model Number |
72404135 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
10/28/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72400023, serial number: n/a , batch/lot number: 1100471676, model/catalog description: balloon fgs 51-60 cm, unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127, serial number: n/a, batch/lot number: 1100776656, model/catalog description: control pump fgs iz, unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.Recurrent incontinence and positional incontinence, and improperly sized cuff are acknowledged within the directions for use (dfu) and are not related to off-label use or failure to follow instructions.A risk review was completed; altered therapeutic response, and size related complications are defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.The reported patient symptom of urinary incontinence is a known risk associated with this device type and indicated as such in the instructions for use.Based on the information available, the conclusion code of known inherent risk of device was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurring incontinence following the initial cuff placement.A cystoscopy was performed, and the urethra did not show any blanching, suggesting that the cuff may be too loose.A surgical procedure was performed to downsize the cuff.No further patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurring incontinence following the initial cuff placement.A cystoscopy was performed, and the urethra did not show any blanching, suggesting that the cuff may be too loose.A surgical procedure was performed to downsize the cuff.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to field b1: adverse event/product problem.Model number/catalog number: 72400023,
serial number: n/a ,
batch/lot number: 1100471676,
model/catalog description: balloon fgs 51-60 cm,
unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127,
serial number: n/a,
batch/lot number: 1100776656,
model/catalog description: control pump fgs iz,
unique identifier (udi) #: (b)(4).Upon receipt at our quality assurance laboratory, the cuff of this artificial urinary sphincter (aus) underwent a thorough analysis.The cuff was visually inspected and tested for leaks.The cuff passed visual inspection.No damage or abnormalities were found.No leaks were found in the cuff.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The device instructions for use (ifu) was reviewed.The reported patient symptom of urinary incontinence was found to be listed in the ifu.A risk review was completed and confirmed that the events of inadequate cuff size/loose cuff and incontinence were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, the reason for cuff - length too long was likely due to the surgeon choosing the wrong length as it was reported that the urethra did not show any blanching, suggesting that the cuff may be too loose.Based on the information available, a conclusion code of unintended use error caused or contributed to even was assigned to this investigation.
|
| |
|