The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 24-apr-2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
It was reported via (b)(4) fda medwatch / fda user facility reports the following information was provided: "corflo ng [nasogastric] tubing with guidewire stylets left post ng placement.Procedure is to place the tube using the guidewire, send patients to x-ray to confirm placement, then remove stylet before using the tube.The tube was utilized for feedings/meds for several days before realizing the guidewire remained.Because the stylet and the ports are the same color, nursing was not alerted to the fact that the stylet had not been removed post imaging.Design flaw.Remove the stylet after confirming placement.Device was hard to use.".
|