| Model Number |
M00558350 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was attempted to be used in the esophagus during an esophagogastroduodenoscopy (egd) procedure to treat stricture performed on (b)(6) 2026.During the procedure, the balloon burst at 8 atm during inflation to the highest rated pressure.The rupture appeared to occur at the junction where the balloon attaches to the catheter.It was reported that no piece of the balloon detached inside the patient.The procedure was completed with another cre fixed wire dilatation balloon device.There were no patient complications have been reported as a result of this event.The patient condition at the conclusion of the procedure was reported to have full recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.Block h11: investigation results the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.With all available information, boston scientific concludes the reported event of balloon burst could not be confirmed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.
|
| |
|
Search Alerts/Recalls
|