• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NORIS MEDICAL LTD TUFF UNICON IMPLANT; TUFF UNICON DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NORIS MEDICAL LTD TUFF UNICON IMPLANT; TUFF UNICON DENTAL IMPLANT Back to Search Results
Model Number NMCF5508
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 02/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2025.On (b)(6) 2026, loosening of implant not related to bone ingrowth was verified.According to the information, the patient has hypertension, periodontitis.At the event, the patient experienced: pain.Observed during the event: mobility.The device was forwarded to the manufacturer.No patient operative or post-operative complications reported.
 
Additional Manufacturer Narrative
The dhr was reviewed for lot # 9001523 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.To date, items of the same lot have been sold without any similar reported event during the period of review.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFF UNICON IMPLANT
Common Device Name
TUFF UNICON DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
Manufacturer (Section G)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS   3688808
Manufacturer Contact
anna dvinska
hataasia 8
nesher, 36888-08
IS   3688808
MDR Report Key24993442
Report Number3011390931-2026-00119
Device Sequence Number15849860
Product Code DZE
UDI-Device Identifier07290116154732
UDI-Public07290116154732
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K240202
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberNMCF5508
Device Catalogue NumberNMCF5508
Device Lot Number9001523
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2026
Initial Report FDA Received Date04/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
Patient Weight111 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-