• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIBUYA (SBY) NIPRO SURDIAL DX HEMODIALYSIS SYSTEM; DIALYSIS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIBUYA (SBY) NIPRO SURDIAL DX HEMODIALYSIS SYSTEM; DIALYSIS MACHINE Back to Search Results
Model Number MC+SDX01
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
Reported condition: loose screw on blood pump rotors on both machines reported by staff.(b)(6): existing blood pump rotor requires replacement.Screws on face plate are loose.Installed new blood pump rotor assembly and performed blood pump occlusion adjustment procedures.The machine operates per the manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO SURDIAL DX HEMODIALYSIS SYSTEM
Common Device Name
DIALYSIS MACHINE
Manufacturer (Section D)
SHIBUYA (SBY)
2-72, wakamiya
kanazawa, isikawa 920-0 054
JA  920-0054
MDR Report Key25001031
Report Number1056186-2026-00016
Device Sequence Number9756447
Product Code KDI
UDI-Device Identifier00383790008815
UDI-Public00383790008815
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/27/2026
Report Date (Section F) 04/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC+SDX01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/07/2025
Event Location Outpatient Treatment Facility
Date Report to Manufacturer04/27/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/27/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Date Report Sent to FDA04/27/2026
-
-