Internal studies concluded that the source of cycle number delay can be traced to metal leaching into the lysis solution from the lysis transfer pump in the alinity m system.Impact on product functionality: based on the data summarized in 616-risk, a potential hazard of delay of results was identified for the scenario in which metal leaching in the lysis buffer can lead to invalidated hpv and sti positive assay controls.Without valid controls, the customer will experience a delay in obtaining valid sample results until the troubleshooting is performed to remove the leached metal in the lysis buffer on the instrument.Conclusion: based on the data, the potential hazard associated with this issue is delay in results for the hr hpv and sti assays.There is no potential for incorrect results for the hpv or sti assays.A decision was made to issue a customer letter to notify customers of the reported issue.This field action, fa-am-sep2024-300 was issued on september 12, 2024.Additionally, this action was reported per 21 cfr806 to fda on september 17, 2024 and is recommended to be reported as an fsca (field safety corrective action).For the alinity m hpv assay, complaints associated with metal leaching causing invalidated hpv positive assay controls may be associated with an unacceptable delay in results, which is associated with a minimal severity.For the alinity m sti assay, complaints associated with meal leaching causing invalidated sti positive assay controls, may be associated with an unacceptable delay in results, which is associated with a major severity.See list below for complete list of additional mdrs associated with fa-am-sep2024-300: 3005248192-2024-00148, 3005248192-2024-00149, 3005248192-2024-00150, 3005248192-2024-00151, 3005248192-2024-00152, 3005248192-2024-00153, 3005248192-2024-00154, 3005248192-2024-00155, 3005248192-2024-00167, 3005248192-2024-00168, 3005248192-2024-00169, 3005248192-2024-00172, 3005248192-2024-00173, 3005248192-2024-00175, 3005248192-2024-00176, 3005248192-2024-00177, 3005248192-2024-00178, 3005248192-2024-00186, 3005248192-2024-00200, 3005248192-2024-00221, 3005248192-2024-00230, 3005248192-2025-00039, 3005248192-2025-00041, 3005248192-2025-00062, 3005248192-2025-00063, 3005248192-2025-00133, 3005248192-2025-00135, 3005248192-2025-00136, 3005248192-2025-00137, 3005248192-2025-00138, 3005248192-2025-00156, 3005248192-2025-00210, 3005248192-2025-00211, 3005248192-2025-00212, 3005248192-2025-00215, 3005248192-2025-00239, 3005248192-2025-00258, 3005248192-2025-00325, 3005248192-2025-00338, 3005248192-2025-00491, 3005248192-2026-00001, 3005248192-2026-00002, 3005248192-2026-00014, 3005248192-2026-00015, 3005248192-2026-00020, 3005248192-2026-00038.
|