| Catalog Number |
GST45T |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/27/2026.D4: batch#: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a lobectomy procedure, it was observed that the staples were malformed after firing.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/11/2026 d4: batch #660d34 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst45t reload was received unfired.In addition, an opened package was received along with the device.The returned reload was loaded into a test device, tested for functionality in the straight position, and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as no malformed staples were observed after fired and the reload was returned without detectable damage.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.Based on the results of this investigation , it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.A manufacturing record evaluation was performed for the finished device batch number 660d34, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|