• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG COAGULATING AND DISSECTING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG COAGULATING AND DISSECTING ELECTRODE Back to Search Results
Model Number 26775UF
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.The device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was reported that "the needle holder does not grip properly, and the electrode has a damaged insulated section." no negative impact in state of health reported.
 
Additional Manufacturer Narrative
Manufacturer's investigation results: the complained product was not returned to karl storz for technical evaluation; therefore, a physical inspection could not be performed.As no investigation report was provided and the device was not available for analysis, a detailed classification of the complaint is not possible.According to the customer's description, the electrode presented a damaged insulation section.Based on the information available and on experience with similar complaints, the observed condition may be associated with mechanical damage occurring during handling and use, such as contact with sharp objects or damage during insertion or removal through a trocar.However, this assumption cannot be confirmed without examination of the affected product.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGULATING AND DISSECTING ELECTRODE
Common Device Name
COAGULATING AND DISSECTING ELECTRODE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key25001752
Report Number9610617-2026-00927
Device Sequence Number18581135
Product Code GEI
UDI-Device Identifier04048551072417
UDI-Public04048551072417
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K944796
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number26775UF
Device Catalogue Number26775UF
Device Lot NumberWM04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2026
Supplement Date Received by Manufacturer05/04/2026
Initial Report FDA Received Date04/27/2026
Supplement Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-