Investigation evaluation: the product said to be involved was returned in a white plastic bag.Provided with the return was an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report due to a pinhole in the proximal portion of the balloon material.The device was returned with the balloon protector placed over the balloon.Prior to functional testing, the balloon protector was removed.During functional testing a cook dilation syringe (ds-60cc-s) from our shelf stock was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was attempted to be inflated.The balloon would not fully inflate or hold pressure, and a leakage was observed from a pinhole in the proximal portion of the balloon material.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the returned device confirmed the report due to a pinhole in the balloon material.A definitive cause for the reported observation could not be determined.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.A leakage/damage to the balloon material can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Another possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all eclipse ttc wire guided balloon dilators are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
|