| Model Number |
V+O+C+S+N+PRO, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issues of it measuring lower peep (positive end expiratory pressure) than set and displaying a patient circuit disconnect alarm was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issues.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ift.
|
| |
|
Event or Problem Description
|
|
It was reported that the device needed service.Upon evaluation of the device, ventec observed that the ventilator measured lower peep (positive end expiratory pressure) than set and was displaying a patient circuit disconnect alarm.There were no reports of patient involvement associated with the reported issues.
|
| |
|
Search Alerts/Recalls
|