| Model Number |
0272 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product remains implanted and in service; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a high out of range shocking impedance measurement.Technical services (ts) was consulted and indicated that the overall shocking impedance trend was stable, although variable values around 120 ohms were noted.No noise events were observed from reviewed episodes.Ts recommended performing isometric testing and suggested increasing the out of range upper limit to 150 ohms.No adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Search Alerts/Recalls
|