SMITH & NEPHEW, INC. LGN REV TIBIA BASE SZ 4 RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
71424014 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication report that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, six years after a right tka surgery was performed, loosening of the tibial baseplate was identified through radiographic imaging performed to assess bone quality.The adverse event was addressed by revision surgery on (b)(6) 2026, during which a lgn rev tibia base sz 4 rt, a lgn cement stem10mmx120mm strt and a lgn ps high flex xlpe sz 3-4 11mm were revised.Intraoperatively, tibial baseplate loosening was confirmed.No signs of infection, osteolysis, or bone loss were observed during the revision procedure.Current health status of patient is unknown.
|
| |
|
Search Alerts/Recalls
|
|
|