• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENTERING; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENTERING; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Catalog Number 103544000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 04/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had left hip dislocation.Previous infected hip.Explanted stem, bipolar head and 28+6 srom head.Implanted srom stem went up to a +12.Downsized bipolar head to 43x28.Implanted 28+3 srom head.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d10 (concomitant).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENTERING
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key25013341
Report Number1818910-2026-07706
Device Sequence Number10295116
Product Code KWY
UDI-Device Identifier10603295003342
UDI-Public(01)10603295003342(17)301130(10)D25112384
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K812672
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number103544000
Device Lot NumberD25112384
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/13/2026
Supplement Date Received by Manufacturer05/07/2026
Initial Report FDA Received Date04/28/2026
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
S-ROM*HEAD FEMORAL COCR 28+6; SROM*STM ST,30+4L NK,18X13X160
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-