|
H3: product analysis :evaluation determined that it is difficult to turn the ring of the attachment, the dial is stuck it is difficult to retract or extend the tube.The likely cause was identified as due to distal bearing is detached.It was also noted that laser markings are faded.B3: date of incident or estimated date of incident not provided to manufacturer.Aware date used as date of incident.G3: aware date used as date of analysis performed as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|