• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NORIS MEDICAL LTD PTERYFIT IMPLANT; PTERYFIT DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NORIS MEDICAL LTD PTERYFIT IMPLANT; PTERYFIT DENTAL IMPLANT Back to Search Results
Model Number NMAF4218
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377)
Date of Event 06/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dhr was reviewed for lot#: 9000955 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.The investigation for this event has not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
The healthcare professional reports the implant was inserted on (b)(6) 2025 in the patient's mouth.On (b)(6) 2025, loss of osseointegration was reported.According to the information, the patient is a smoker.Observed during the event: bone loss/mobility.The device was forwarded to the manufacturer.No patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PTERYFIT IMPLANT
Common Device Name
PTERYFIT DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
Manufacturer (Section G)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS   3688808
Manufacturer Contact
anna dvinska
hataasia 8
nesher, 36888-08
IS   3688808
MDR Report Key25015697
Report Number3011390931-2026-00131
Device Sequence Number9542355
Product Code DZE
UDI-Device Identifier07290108817102
UDI-Public07290108817102
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberNMAF4218
Device Catalogue NumberNMAF4218
Device Lot Number9000955
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2026
Initial Report FDA Received Date04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
Patient Weight50 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-