MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404013
Medical Device Problem Codes Collapse (1099); Inflation Problem (1310); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the cylinders of an inflatable penile prosthesis (ipp) could not be inflated.Air was observed within the device, and the pump remained flat.Subsequently, a surgical procedure was performed to remove and replace the ipp.No complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1000076790.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404161.Serial number: n/a.Batch/lot number: 0176512005.Model/catalog description: reservoir 65ml pc.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Incomplete inflation and air in the system are acknowledged within the directions for use (dfu) and are not related to off-label use or failure to follow instructions.A risk review was completed, and inflation issue and air in the system are defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25015871
Report Number2124215-2026-22697
Device Sequence Number18555538
Product Code JCW
UDI-Device Identifier00878953002705
UDI-Public(01)00878953002705(17)220807(10)183326010
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/07/2022
Device Model Number72404013
Device Catalogue Number72404013
Device Lot Number0183326010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2026
Initial Report FDA Received Date04/28/2026
Date Device Manufactured08/07/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age36 YR
Patient SexMale
Patient Weight69 KG
-
-