| Catalog Number |
21-7600-24 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.B3 - date of event: since the exact event date was unknown, it was updated as first day of the month of awareness (mar-2026).
|
| |
|
Event or Problem Description
|
|
It was reported that there was liquid leakage from the connection part of the syringe during priming.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
D3 - postal code, h3, h6: updated.The suspected product was returned for evaluation.Visual inspection was performed and was confirmed that there were no anomalies noted.The lot history review confirmed that there were no previous conformities noted.The tube luer bond junction of the cassette was leak tested using hydrostatic pressure pump and a leak was confirmed.The luer tube bond of the leaking sets was separated and the tube and bond pocket was inspected.A solvent channel was observed on the tube.The allegation made by the complainant was confirmed.The probable root cause of the event was due to leakage at luer tube bond.
|
| |
|