• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. NRFIT CADD 100ML EPID CASSETTES; SET, IV, FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. NRFIT CADD 100ML EPID CASSETTES; SET, IV, FLUID TRANSFER Back to Search Results
Catalog Number 21-7600-24
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.B3 - date of event: since the exact event date was unknown, it was updated as first day of the month of awareness (mar-2026).
 
Event or Problem Description
It was reported that there was liquid leakage from the connection part of the syringe during priming.There was no patient involvement.
 
Additional Manufacturer Narrative
D3 - postal code, h3, h6: updated.The suspected product was returned for evaluation.Visual inspection was performed and was confirmed that there were no anomalies noted.The lot history review confirmed that there were no previous conformities noted.The tube luer bond junction of the cassette was leak tested using hydrostatic pressure pump and a leak was confirmed.The allegation made by the complainant was confirmed.The probable root cause of the event was due to leakage at luer tube bond.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NRFIT CADD 100ML EPID CASSETTES
Common Device Name
SET, IV, FLUID TRANSFER
Manufacturer (Section D)
ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key25016601
Report Number3012307300-2026-04279
Device Sequence Number16035510
Product Code LHI
UDI-Device Identifier10610586044007
UDI-Public(01)10610586044007(17)270106(10)6082332
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K162219
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21-7600-24
Device Lot Number6082332
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/23/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2026
Supplement Date Received by Manufacturer06/05/2026
Initial Report FDA Received Date04/28/2026
Supplement Report FDA Received Date06/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-