• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TUFTEX EMBOLECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TUFTEX EMBOLECTOMY CATHETER Back to Search Results
Catalog Number 1601-44
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an embolectomy procedure, the tuftex embolectomy catheter balloon ruptured while removing a clot.The device had been checked prior to use and was confirmed to be acceptable, with successful inflation and no leakage observed.The saline volume used for balloon inflation did not exceed the recommended limit.There was no indication of abnormal storage conditions.The device was reportedly used in a stenotic region.No additional incision was made, and no sharp object was used on the balloon.No catheter fragments were retained in the patient.The procedure was successfully completed using a replacement device.No patient injury or adverse impact to the patient's health was reported.
 
Additional Manufacturer Narrative
The device has not been received for investigation.Therefore, we could not conclusively determine the root cause of the reported incident.It is possible that procedural factors/anatomical features such as calcification, or plaque caused/contributed to the reported issue.As stated in the instructions for use (ifu): "in common with other catheterization procedures, complications may occur.These may include intimal disruption, vessel wall perforation, balloon rupture with fragmentation, tip separation, local or systemic infection, arterial thrombosis, local hematomas, air embolus, rupture of aneurysm, arterial spasm, distal embolus of blood clots or arteriosclerotic plaque, arterial dissection, and hemorrhage." the production and traceability record for the device was reviewed; no issues were found during manufacturing or packaging that would cause or contribute to the reported event.All quality control tests were completed successfully and met specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFTEX EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key25016662
Report Number1220948-2026-00099
Device Sequence Number21208838
Product Code DXE
UDI-Device Identifier00840663100484
UDI-Public(01)00840663100484(11)210812(17)270728(10)SLC5499(91)1601-44
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1601-44
Device Lot NumberSLC5499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2026
Initial Report FDA Received Date04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-