| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause cannot be identified.Based on the results of the investigation, it is likely that the following led to the malfunction: degradation problem.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
The bronchovideoscope was returned to the service center with a report of failed leak test at distal tip. this does not indicate an mdr reportable event. however, mdr reportable failure was found during testing at the service center.During inspection of the device, corrosion inside lens and fiber was found.There was no patient involvement.
|
| |
|