|
During validation after installation, the customer experienced intermittent issues with ih-cell a1, b reagent recognition (barcode unreadable / reagent empty despite full vial).Troubleshooting confirmed correct reagent reference and stable connectivity.A subsequent technical support investigation identified incorrect reagent position mapping (positions 5 and 6 mapped to the same location), leading to aspiration attempts from an unloaded position.A reagent area teaching procedure corrected the mapping.Problem: the instrument encountered a target out of range error 0x030402 for the pipettor y axis (see qn (b)(4)), action: the fse adjusted the positions to overcome this error.The fse did not follow correctly the "ih-500 ni pipettor object out of range" procedure (*), combined with the lack of the software safety feature the wrong position was saved.Consequently.Position 3 was accessed instead of position 4.(*) most probably did not drive the pipettor to the corresponding position in the 2nd row, but a position beside the affected position (see 6.2.4 in the procedure).Testing: the issue was detected during a control run.After replacing the instrument parameter with an older version, the fse was able to setup the reagent positions correctly.Injury: no injury or harm reported, malfunction detected by control run, log analysis confirmed that no patient sample were processed in the period were the reagent coordínate were not setup incorrectly.
|