Model Number 3401
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 04/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this subcutaneous implantable cardioverter defibrillator (s-icd) system presented to the emergency room (er).Upon review, it was noted that this s-icd system delivered an inappropriate electric shock due to oversensing and noise.Low amplitude was also noted.Isometric testing was performed and the noise was unable to reproduce.Further isometric testing and a chest x-ray were recommended to rule out lead integrity.The representative reprogrammed to pri and discharged the patient.Then, a revision procedure was performed and the s-icd system was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
OSCOR INC
3816 desoto blvd.
palm harbor FL 34683
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
saint paul MN 55112
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key25023057
Report Number2124215-2026-22991
Device Sequence Number18383526
Product Code LWS
UDI-Device Identifier00802526575600
UDI-Public(01)00802526575600(17)171101
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/01/2017
Device Model Number3401
Device Catalogue Number3401
Device Lot NumberA129092
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2026
Supplement Date Received by Manufacturer06/23/2026
Initial Report FDA Received Date04/28/2026
Supplement Report FDA Received Date07/01/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age32 YR
Patient SexMale
-
-