| Model Number |
V+PRO, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the device was not returned to ventec for an evaluation.The device was evaluated by an authorized service provider (asp) where the reported issue of it providing low fio2 (fraction of inspired oxygen) readings was initially confirmed, however, after subsequent testing of the device the reported issue could no longer be duplicated.Proper device operation was confirmed through functional and performance testing.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
An authorized service provider (asp) contacted ventec to report that the device was providing low fio2 (fraction of inspired oxygen) readings.There were no reports of patient involvement associated with the reported event.
|
| |
|