| Model Number |
480440-06 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Material Fragmentation (1261); Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive received the synchroseal to perform failure analysis.Visual inspection found the cut electrode thermally damaged on the jaws.The cut electrode was thermally damaged, and there may be missing pieces; however, the size of the missing pieces cannot be confirmed, indicating that a fragment may have detached from the instrument.The synchroseal instrument was placed on in house system and passed engagement and recognition tests.The seal test was performed and passed all three attempts.The probable root cause may be attributed to either damage during use or the manufacturing process.Damage during use can result from instrument collisions or improper intraoperative cleaning with another instrument.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted hysterectomy ¿ benign surgical procedure, three synchrseal within one surgery were defective.After one or two uses, the coating in the jaws came loose and the device displayed an error message.The instrument then became unusable.Repeated error 25913 occured mid procedure.The nurse replaced the syncroseal instrument with a backup synchroseal instrument and the error reoccurred.The clinical sales representative (csr) power cycled e-100 generator, the issue appeared to occur less but it was still occurring.The third synchroseal instrument that was used had the same error occur.The error message displayed that homeostasis was compromised and the sealing process was not successful.Error logs show error 25913 - e-100 error: recoverable error: instrument cut electrodes shorted (p1:0x10000001) / rec-01.The csr stated the instrument branches were not too dirty, not carbonized.The instrument was cleaned and reinserted, but the error persisted.Afterwards, the csr stated that the plastic on the synchroseal instrument appeared as if it was liquified and/or some parts were missing.At the time, it was not clear yet if parts fell into patient.The tse requested photos of the broken instrument including serial numbers and will add the pictures once they arrive.At the time of the call, it was not clear if all instruments were from the same charge of production.The field service engineer (fse) was informed to replace e-100 generator.The case was cloned to the customer support queue to ensure proper handling of incident.The procedure would be completed without bipolar energy from the synchroseal instrument.The customer would not use e-100 generator until clarification.The monopolar energy from erbe generator would be used instead (also different instruments).No fragment fell into the patient.The user completed the procedure and no known impact or patient consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did not receive the da vinci product with an alleged issue to perform failure analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the da vinci product with an alleged issue to perform failure analysis.Upon visual inspection, no issues were found that would be related to the reported event.The e100 generator was installed onto a golden system where it was tested with 10 power cycles and 50 energy cycle cut/seal actions.Reported complaint could not be replicated.Probable root cause is attributed to defective generator.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Search Alerts/Recalls
|