| Model Number |
6882000 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the replacement of the pressure transducer printed circuit board (pcb) solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.Pressure transducer pcb measures the pressure, conveyed via the pressure tube connected to this block by its differential pressure transducer.With differential reference to the ambient pressure, the output signal is proportional to the measured pressure thus giving a linear measurement in the range -40 cmh2o to +160 cmh2o.A positive end expiratory pressure (peep) is maintained in the alveoli and may prevent the collapse of the airways.The device's logs were not provided for further analysis.The defective part was not returned for investigation, despite request.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to pressure transducer pcb.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator generated an alarm indicating low positive end expiratory pressure (peep).There was no patient harm.Manufacturer's ref.#: (b)(4).
|
| |
|