C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
1806060 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.The investigation is inconclusive for the reported fracture, material separation and migration as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the litigation process that, sometime after the port placement, the patient was diagnosed with a fractured chest port tip located in the right atrium.However, the current status of the patient remains unknown.
|
| |
|
Search Alerts/Recalls
|
|
|