• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. BLAKE; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. BLAKE; CATHETER, IRRIGATION Back to Search Results
Catalog Number 2233
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient underwent an unknown urological surgery on (b)(6) 2026 and drain was used.When the package was opened, it was found that the drain had been adhered to the needle.When they were separated, it was found that there was damage to the drain part.Another product was used to complete the case.There were no adverse consequences to the patient.Further details are not provided.Upon receipt and analysis, drain had chip-off marks.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) if further details are received at a later date a supplemental medwatch will be sent.H3 evaluation: complaint sample review: one complaint sample of drain with trocar was received for evaluation, product was checked visually and found that there was no sticking mark visible on the trocar, but the drain had chip-off marks at the location.As per standard practice, 100% functional test and 100% visual inspection was carried out, before and after packing of finished goods, prior to the product release.There was no scope to miss such defect, at manufacturing / release stage.Documents review: during investigation of complaint, batch review of in-process and final packaging inspections, as well as the manufacturing process is performed, these products are reported to be manufactured, inspected, and packaged in conformance with customer's specifications and have been found to be acceptable within all parameters.No discrepancy as per the reported defect is observed.Retention sample review: no negative observation was found.Review of defective sample's image / video: not applicable, as there was no complaint sample image / video received for evaluation.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAKE
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
DEGANIA INDIA
251, sector-6, imt manesar
gurugram 12205 2
IN   122052
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25027777
Report Number2210968-2026-04627
Device Sequence Number8992096
Product Code GBX
UDI-Device Identifier10705031003613
UDI-Public(01)10705031003613(17)291230(10)J2403061
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
CL I EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2233
Device Lot NumberJ2403061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-