• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES; TUBES, VACUUM SAMPLE, WITH ANTICOAGULANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES; TUBES, VACUUM SAMPLE, WITH ANTICOAGULANT Back to Search Results
Catalog Number 367861
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k213670;k231373 there were multiple medical device types reported to be involved.The information for the additional device type is as follows d.2.B medical device type: jka d.2.A common device name: tubes, vials, systems, serum separators, blood collection e.1.Initial reporter facility name: (b)(6).Investigation summary: bd received 5 photos for investigation which showed white streaks in the sidewall and bottom of the tube.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.This complaint has been confirmed for the indicated failure mode: molding defect.No definitive potential cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported while using bd vacutainer® k2 edta (k2e) 7.2mg blood collection tubes, foreign matter was seen inside of one tube.No adverse impact was reported regarding the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES
Common Device Name
TUBES, VACUUM SAMPLE, WITH ANTICOAGULANT
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key25029594
Report Number1917413-2026-00340
Device Sequence Number10126960
Product Code GIM
UDI-Device Identifier30382903678618
UDI-Public(01)30382903678618(11)250716(17)261130(10)5197583
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
BK230980
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number367861
Device Lot Number5197583
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-