SMITH & NEPHEW, INC. Q-FIX KNTLS ASA 1.8MM W/1 UB SUT BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72205882 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Codes |
Perforation (2001); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, after the q-fix anchor was deployed according to the surgical technique, the mechanism failed and the repair suture never made it through the chinese finger mechanism when shuttle the repair suture, rendering it unusable.The procedure was after a non-significant delay, using an equivalent s+n back up product on a new drilled bone hole, leaving a void but still filled with the suture ball of the anchor that failed.The patient is recovering well and no further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|