Model Number BF-Q190
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely the following led to the malfunction: component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The bronchovideoscope was returned to the service center with a report of dent on the inside of channel and has scratches.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, objective lens glue is chipped was found.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25031825
Report Number9610595-2026-34884
Device Sequence Number13295352
Product Code EOQ
UDI-Device Identifier04953170335198
UDI-Public04953170335198
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K121959
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-Q190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-