| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
| Date of Event |
04/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint revealed that the patient wore the zio at for 3 of the 14 days prescribed.The patient does not have sensitive skin, however they do have a known history of reaction to medical adhesive.The exact cause of the reported event cannot be determined; however, the patient¿s preexisting history cannot be ruled out as a contributory factor.Skin irritation is a known inherent risk of the device.The device manual warnings section read as follows: do not use the zio at patch on patients with known allergic reaction to adhesives or hydrogels or with a family history of adhesive skin allergies.Patient may experience skin irritation.If skin irritation such as severe redness, itching or allergic symptoms develop, remove the zio at patch from the patient¿s chest.This event is being reported per 21cfr803 as a serious injury.This report does not constitute an admission by rhythm that the product described in this report has any defects, or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
|
| |
|
Event or Problem Description
|
|
The patient reported seeking evaluation from a healthcare provider for itching, redness, pain, burning sensation and a rash at the zio at device application site.The device was removed.According to the patient, the hcp prescribed an unspecified medication.
|
| |
|