| Catalog Number |
ECH60L |
| Medical Device Problem Codes |
Failure to Cut (2587); Difficult to Open or Close (2921); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/29/2026.D4: batch #unk.Additional information was requested, and the following was obtained: no patient consequences.Device discarded.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: was the firing sequence started but not completed? if yes: did the device deliver any staples? if yes, were the staples formed properly? if yes, was the staple line complete? did the device cut? if yes, was the cut line complete? if yes, was there any issue with the cut line? was there any difficulty opening the device jaws? a manufacturing record evaluation was performed for the finished #ech60l, with lot/batch #a97y1m, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure, it was observed that the stapler became stuck at the end of the section.An alternative method was used to open the stapler and release the tissue.There was no patient consequence nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/28/2026.Additional information was requested, and the following was obtained: was the firing sequence started but not completed? if yes: did the device deliver any staples? if yes, were the staples formed properly? if yes, was the staple line complete? did the device cut? if yes, was the cut line complete? if yes, was there any issue with the cut line? was there any difficulty opening the device jaws? the device delivered all staples correctly and completely.There was a problem with the cutting line at the end of the zone, the surgeon managed to open the jaws to release the tissue.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: did the device cut? if yes, was the cut line complete? if yes, was there any issue with the cut line such as jagged, dull knife, irregular, etc?.
|
| |
|
Search Alerts/Recalls
|
|
|