Model Number 3225
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 04/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient experienced ineffective therapy.Diagnostics showed high impedances.As a result, the lead was explanted and replaced to address the issue.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCLAIM LEAD, 90 CM
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key25033478
Report Number1627487-2026-02071
Device Sequence Number9176309
Product Code LGW
UDI-Device Identifier05415067017758
UDI-Public(01)05415067017758(10)5414863(17)180228
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Model Number3225
Device Catalogue Number3225
Device Lot Number5414863
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/10/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS IPG
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexMale
Patient Weight100 KG
-
-