Catalog Number ECH60C
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 4/29/2026.B3: unknown.D4: batch #a98h8v.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech60c device was returned with no apparent damage and with no reload present.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb was noted to be non-functional.No conclusion could be reached as to what may have caused the pcb to be damaged.The device event log was reviewed.A series of events were found that may indicate intermittent power issues.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number a98h8v, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown thoracic procedure, it was observed that the battery did not operate.Another device was used before loading the cartridge to complete the case.There were no adverse consequences to the patient.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key25033518
Report Number3005075853-2026-03235
Device Sequence Number9039390
Product Code GDW
UDI-Device Identifier10705036034971
UDI-Public(01)10705036034971(17)280731(10)A98H8V
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K213633
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECH60C
Device Lot NumberA98H8V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-