| Catalog Number |
242274 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary = > the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: - other damages caused by customer distal tip damaged - laser welding damaged, needle - optical system, optical components broken lenses in optical system - cosmetic issue scratches/dents rust/discoloration visual: deformed/bent, broken (2+ pieces): chipped/shaved/delaminated, visual: scratched/nicked.Per service report, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review = > a manufacturing record evaluation was performed for the finished device serial number (b)(6), and no non-conformance was identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during in-house engineering evaluation, it was determined that the hd c-mnt scp,1.9,30,60,mitek device was chipped/shaved/delaminated.It was initially reported that the tip of the scope was bent.The device was not used in surgery.There was no delay in the surgery.The fms vue ii was swapped with a backup pump.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|