MEDTRONIC IRELAND EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVOLUTFX-34
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a transcatheter aortic valve implant procedure, a pre-implant balloon aortic valvuloplasty (bav) was com pleted due to it being standard for the implanting physician.The dcs was then inserted into the patient, and the valve was advanced to the aortic annulus.The valve was then partially deployed but subsequently recaptured.While recapturing the valve, an infold was noted so the valve was withdrawn from the patient's body.A new dcs and valve were then prepared and inserted into the patient's body.The valve was then deployed at the aortic annulus but subsequently was recaptured.Upon second deployment, an infold was noted so the valve was recaptured and withdrawn from the patient's body.A third valve was then prepared along with a new dcs.The dcs was then inserted into the patient's body and valve was successfully implanted.A 20-minute procedural delay occurred due to the valve infolds.Additional information was received that there was no issue during the loading of both valves.The first valve was recaptured because the valve dislodged into the sinus while attached to the dcs at 80% deployment.The second valve was recaptured because the valve was placed at 0 mm on the left coronary cusp (lcc) so the physician felt the valve depth was too high.
 
Additional Manufacturer Narrative
D10.Continuation - concomitant medical devices in use during the event include: medtronic product id: evfxplus-34; product serial/lot number: r085769; product type: 0195-heart valves; implant date: na; explant date: na.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received that the deployment starting point was at the mid pigtail.The physician started with a non-medtronic guidewire and then changed to a different non-medtronic guidewire during the procedure.The first valve "popped up" twice during the attempted deployment.After the second recapture, the infold in the valve frame was noticed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX DCS
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key25035147
Report Number9612164-2026-02238
Device Sequence Number9176216
Product Code NPT
UDI-Device Identifier00763000942717
UDI-Public(01)00763000942717(17)270918(10)0013028595(20)01
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD-EVOLUTFX-34
Device Catalogue NumberD-EVOLUTFX-34
Device Lot Number0013028595
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date04/29/2026
Supplement Report FDA Received Date05/27/2026
Date Device Manufactured09/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-