| Model Number |
2C8519 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: h4, h6 (update codes), h11.H11: the actual device was not available; however, two photographs of the sample were provided for evaluation.Visual inspection was performed which showed that the tubing was separated from the second y-site clearlink in the downstream part.The reported condition was verified.The cause of the issue was related to manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a continu-flo solution set tubing separated at the clearlink y-site.The event occurred while unscrewing the luer lock from the iv line in the icu.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|